Tech Specs:
Ease-of-use
- Simple,
easy-to-use, one-button operation.
- More
instructive voice prompts guide user through rescue.
- Outstanding
voice prompt quality and clarity.
- Lighter
weight in a new user-friendly form factor.
Features & Benefits
is patented self-testing technology determines whether an AED is in working order
and ready for a rescue. If a Cardiac Science AED fails any one of its self-tests, an
audible alarm and accompanying visual cue (a green/red status indicator) provides an
effective alert to users.
-
COMPREHENSIVE TESTING PROTOCOL
Unlike some other manufacturers, Cardiac Science tests AEDs with a comprehensive
series of analyses. All testing data is stored within the device.
• Daily Tests: Battery; Pads for both presence and function; internal electronics
• Weekly: Partial energy charge; software
• Monthly: Internal electronics; full energy simulated rescue
Powerheart AEDs use pre-connected pads that can be used in either position on a
patient. is means that either pad can be placed in the upper-right or lower-left chest
position making pad placement much easier than other manufacturer’s pads.
-
VARIABLE ESCALATING ENERGY
Cardiac Science AEDs analyze the patient using our proprietary RhythmX technology
to determine his/her electrical impedance (resistance levels) and then customize an
appropriate energy level for that individual patient. And, if more than one shock is
necessary to convert the patient, our analysis software utilizes variable escalating
energy to deliver therapy at an appropriate higher level.
|