Every day you save lives, and every day Philips HeartStart MRx and Connected Care Solutions offer advances that matter to you. HeartStart MRx has always been packed with innovative functionality. As the first advanced networked monitor/defibrillator, HeartStart MRx incorporates many of the same algorithms, patient connections, and configuration choices as Philips premier bedside monitors.
Product Specifications
Physical
|
Defibrillator Model |
HeartStart MRx (M3536A) |
|
Dimensions |
Without external paddles: 12.4 in. (W) x 7.7 in. (D) x 11.7 in. (H) (313 mm x 195 mm x 295 mm). With external paddles: 12.4 in. (W) x 7.7 in. (D) x 13.4 in. (H) (313 mm x 195 mm x 340 mm) |
|
Weight |
13.2 lbs. (6 kg): base unit with 1 battery, pads and pads cable. 13 lbs. (5.9 kg) with optional 75 mm strip chart printer. Paddle tray and external standard paddles add less than 2.5 lbs. (1.1 kg). Carrying case adds 4.1 lbs. (1.86 kg). |
Environmental and physical requirements
|
Solids/Water Resistance |
IP24 |
|
Temperature |
Operating: 32º - 113º F (0º - 45º C) Storage: -4º - 158º F (-20º - 70º C) |
|
Humidity |
Operating: 0% to 95% relative |
|
Altitude |
Operating: 0 to 15,000 ft (0 to 4,500 m) Storage: 0 to 15,000 ft (0 to 4,500 m) |
|
Mechanical Shock |
Bump: IEC 68-2-29 Freefall: IEC 68-2-32 |
|
Vibration |
Operating: MIL STD 810E 514.4 Category 6 Helicopter, General Storage, UH60 Non-Operating: IEC 68-2-6 Swept Sine Vibration and IEC 68-2-64 Random Vibration |
|
Safety |
Meets EN 60601-1, UL 2601-1, CSA C22.2 No. 601-1 |
Display
|
Dimensions |
8.4’’ diagonal (128 mm x 171 mm) |
|
Type |
TFT color LCD |
|
Resolution |
480 x 640 pixels (VGA) |
|
Wave Viewing Time |
5 seconds (ECG) |
Defibrillation
|
Waveform |
Truncated Exponential Biphasic.Waveform parameters adjusted as a function of patient impedance. |
|
Output Energy |
Manual (selected): 1-10, 15, 20, 30, 50, 70, 100, 120, 150, 170, 200 Joules into a 50 Ohm load AED Mode (single energy output): 150 Joules into a 50 ohm load. |
|
Charge Time |
Less than 5 seconds to 200 Joules with a new, fully charged lithium ion battery at 25º C |
|
Shock Delivery |
Via multifunction defib electrode pads or paddles |
|
Shock-to-Shock Cycle Time |
Typically less than 20 seconds |
|
Patient Impedance Range |
Minimum: 15 Ohm (internal defibrillation); 25 Ohm (external defibrillation) Maximum: 180 Ohm |
|
AED Mode |
Shock advisory sensitivity and specificity meet AAMI DF-39 guidelines |
Battery
|
Type |
6.3 Ah, 14.8 V, rechargeable lithium ion |
|
Dimensions |
6.5" (H) x 3.8" (W) x 1.6" (D) (165 mm x 95 mm x 42 mm) |
|
Weight |
1.6 lb. (0.73 kg) |
|
Charge Time |
Approximately 3 hours from fully depleted to 100%, 90 minutes from fully depleted to 80% |
|
Capacity |
At least 5 hours of continuous 12-lead ECG, SpO2,and CO2 monitoring, with NBP every 15 minutes on one new, fully charged battery. At least 3.5 hours of continuous 12-lead ECG, SpO2, and CO2 monitoring, with NBP every 15 minutes and pacing at 180 ppm at 160 mA on one new, fully charged battery. |
|
Battery Indicators |
Battery gauge on battery, capacity indicator on display; flashing RFU indicator, chirp, and ‘Low Battery’ message appears on display for low battery condition, when 10 minutes of monitoring time and 6 maximum energy discharges remain (with a new battery at room temperature, 25º C) |
Strip chart recorder
|
Recorder |
Standard: 50 mm (paper width) thermal array printer Optional: 75 mm (paper width) thermal array printer |
|
Continuous ECG Strip |
Prints primary ECG lead with event annotations and measurements in real-time or with 10-second delay |
|
Auto Printing |
Recorder can be configured to print marked events, charge, shock and alarms |
|
Reports |
Event Summary, 12-Lead, Operational Check, Configuration, Status Log, and Device Information |
|
Paper Size |
1.97 in. (50 mm) W by 100 ft. (30 m) L 2.95 in. (75 mm) W by 100 ft. (30 m) L |
Data storage
|
Internal |
12 hours of continuous ECG waveforms and events, plus 50 12-lead ECG reports |
|
Data Card |
12 hours of continuous ECG waveforms and events, plus 50 12-lead ECG reports, on a CompactFlash memory card |
ECG and arrhythmia monitoring
|
Input |
Up to 4 ECG waves displayed and up to 3 ECG waves print simultaneously. Lead I, II, or III obtained through 3-lead ECG cable and separate monitoring electrodes. With 5-lead cable, obtain leads I, II, III, aVR, aVL, aVF, or V. Pads ECG obtained through 2 multifunction defibrillation electrode pads. |
|
Lead Fault |
‘Lead Off ’ message and dashed line displayed, if an electrode or lead wire becomes disconnected |
|
Pads Fault |
Dashed line displayed if a pad becomes disconnected. |
|
Heart Rate Display |
Digital readout on display 15 to 300 bpm, accuracy ±10% |
|
Heart Rate/Arrhythmia Alarms |
HR, Asystole,VFIB/VTACH,VTACH, extreme tachycardia, extreme bradycardia, PVC rate |
|
ECG Size |
2.5, 5,10, 20, 40 mm/mV, autogain |
SpO2 pulse oximetry
|
Range |
0 to 100% |
|
Resolution |
1% |
|
Alarm Range |
Low Limit: 50 to 99% (Adult/Pediatric) High Limit: 51 to 100% (Adult/Pediatric) |
|
Alarm Delay |
10 seconds |
Noninvasive Blood Pressure
|
Pressure Range |
Systolic: 40 to 260 mmHg Diastolic: 20 to 200 mmHg |
|
Initial Pressure |
Adult: 160 mmHg Pediatric: 120 mmHg |
|
Maximum Pressure |
280 mmHg |
|
Alarm Range |
Systolic high limit: 30 - 270 (Adult), 35 - 180 (Pediatric) Systolic low limit: 30 - 265 (Adult), 30 - 175 (Pediatric) Diastolic high limit: 18 - 240 (Adult), 18 - 150 (Pediatric) Diastolic low limit: 10 - 240 (Adult), 10 - 145 (Pediatric) |
End-tidal CO2
|
Range |
0 to 99 mmHg |
|
Resolution |
1 mmHg (0.1 kPa) |
|
Sample Size |
50 ml per minute |
|
Alarm Range |
Low Limit: 10 to 95 mmHg (Adult/Pediatric) High Limit: 20 to 100 mmHg (Adult/Pediatric) |
12-lead ECG
|
Input |
12-Lead cable: leads I, II, III, aVR, aVL, aVF,V/C1-V/C6 |
|
Display View |
All 12-lead ECG waves display simultaneously |
|
Strip Recorder |
All 12-leads print on the strip chart printer in 3x4 format |
|
Transmission |
CompactFlash data card; Bluetooth® dial-up Internet connection, Bluetooth® FTP |
CPR feedback and measurement
|
Compression Depth |
Target: -1.50 to -2.00 in. (-38 to -51 mm) |
|
Compression Rate |
Target: 90 to 120 cpm |
|
Ventilation Volume |
Graphic indicator: empty, 1/3-full, 2/3-full, full |
|
Ventilation Rate |
Target for breaths delivered before 60 seconds since last compression: 6 to 16 vpm Target for breaths delivered beyond 60 seconds since last compression: 9 to 16 vpm |
Noninvasive pacing
|
Waveform |
Monophasic Truncated Exponential |
|
Current Pulse Amplitude |
10 mA to 160 mA (5 mA resolution); accuracy 10 mA to 50 mA ± 5 mA, 50 mA - 160 m A ± 10% |
|
Pulse Width |
40 ms with ± 10% accuracy |
|
Rate |
30 ppm to 180 ppm (10 ppm increments); accuracy ± 1.5% |
|
Modes |
Demand or Fixed Rate |
|
Refractory Period |
340 msec (30 to 80 ppm); 240 msec (90 to 180 ppm) |