Medtronic 5375


CLICK HERE for a larger picture

Specifications


The Medtronic® Model 5375 External demand Pulse Generator is an external, battery-powered single chamber pulse generator designed for atrial or ventricular temporary cardiac pacing.  Its clinical applications include: 1) pre-. intra, and postoperative managment of cardiac surgery patients; 2)short-term treatment of arrhytmias and heart block; and 3) emergency cardiac pacing.   The model 5375 can provide demand (R-wave inhibited) or asynchronous (fixed-rate) pacing and can be used with myocardial or transvenous lead systems in either a bipolar or unipolar configuration.  With its calibrated controls the model 5375 can be used to determine both stimulation thresholds and R-wave potensials sensed by the electrodes of a pacing lead system.

Dimension
1.5 in.(3.8cm)  x  3 in.(7.6 cm)  x  5.7 in.(14.5 cm). 
Weight:  Approximately 476 grams with battery.
Power Supply:
9-volt alkaline battery Eveready Energizer 522**
 OTHER SPECIFICATIONS:
  • Rate Range 30 to 180 pulses per minute continuosly adjustable.
  • Rate Calibration Accurracy: 5% of indicated rate from 50 to 150 ppm;  10% of indicated dial rate outside these limits.
  • Output Current Range... constant current from 0.1 to 20 milliamperes, regardless of lead up to a maximum of 18 volts**
  • Output Current Calibration Accurracy  ± 10% or 0.25 mA (whichever is greater)** from 0.5 mA to 20 mA; accuracy not calibrated below 0.5 mA.
  • Pulse Width: 1.8 milliseconds ± 0.2 milliseconds.
  • Maximum Sensitivity (Sensivity Controll fully clockwise) ... 1.0 millivolts ± 0.5 millivolts for a 40 millisecond, sine-squared pulse.
  • Refractory Period:  250 milliseconds nominal, initiated by the emission of a pacing pulse or the sensing of an R-wave.
** Refer to manual for further information.
Shipping price:  will be according to the bayer location
Our Preferred Method of Payment is paypal. We also accept Visa, MC, Discover, Wire Transfer or Certified Funds.
Guaranty: If the equipment is defective you can contact us within 30 day and we will replace the defective item for you if a replacement is available. If the replacement is not available a refund (not including shipping) will be issued in the form of the original payment method or a company check. Please consider the following terms in such cases:


  • For replacement, merchandise must be returned within 30 days of receipt and in the original packaging accompanied with packing slip.

  • Returned items must not have any signs of abuse or intentional damage. We carefully inspect returned items upon receipt before issuing refunds, credits, or exchanges and compare them with recorded serial numbers.

  • Merchandise should be carefully packed and insured for the full purchase price.

  • Buyer is responsible for all return shipping charges.


The sale of this item may be subject to regulation by the U.S. Food and Drug Administration and state and local regulatory agencies. If so, do not bid on this item unless you are an authorized purchaser. If the item is subject to FDA regulation,our company will  verify your status as an authorized purchaser of this item before shipping the the equipment.